Brands |
N/A Dytran |
Frequency Response |
N/A 1.0 to 10000 Hz |
Type |
N/A Integrated Electronic Piezoelectric (IEPE) |
Housing Material |
N/A Titanium Alloy |
Maximum Weight (Less Cable) |
N/A 0.18 oz5.0 g |
Mounting Provision |
N/A Adhesive |
Element Style Material |
N/A Crystal Shear Mode |
Maximum Peak Shock |
N/A 3,000 g29430 m/s² |
Maximum Peak Vibration (±) |
N/A 1000 g5886 m/s² |
Operating Temperature Range |
N/A -60 to +250 ºF-54 to +121 ºC |
Environmental Seal |
N/A Hermetic |
Sensitivity |
N/A 100 mV/g10.19 mV/m/s² |
Sensitivity Tolerance (±) |
N/A 10 % |
Note for Sensitivity |
N/A Measured at 100 Hz, 1 Grms per ISA RP 37.2. |
Full Scale (F.S.) Range for ± 5 Volts (V) Voltage Output |
N/A 50 g490.00 m/s² |
Frequency Response Tolerance |
N/A 10 % |
Resonant Frequency |
N/A > 30 kHz |
Full Scale (F.S.) Linearity (±) |
N/A 2 % |
Note for Full Scale (F.S.) Linearity (±) |
N/A Percent of full scale or any lesser designated full scale range, zero-based best fit straight line method. |
Root-Mean-Square (RMS) Broadband Resolution |
N/A 0.0004 g0.004 m/s² |
Maximum Transverse Sensitivity |
N/A 5 % |
Maximum Strain Sensitivity |
N/A 0.004 g/µε0.04 m/s²/µε |
Power Supply Current |
N/A 2 to 20 mA |
Note for Power Supply Current |
N/A Power only with Dytran or Dytran approved current source type power unit. Do not supply power without current limiting. You will destroy the integral electronics. This will void the warranty. |
Compliance Voltage Range |
N/A +18 to +30 V |
Direct Current (DC) Operating Bias Voltage |
N/A +11 to +13 V |
Discharge Time Constant |
N/A 0.8 to 2.0 sec |
Nominal Output Impedance |
N/A 100 O |
Output Signal Polarity for Acceleration Toward Top |
N/A Positive |
Minimum Electrical Isolation, Case Ground to Mounting Surface |
N/A 10 MO |
Note |
N/A In the interest of constant product improvement, we reserve the right to change specifications without notice. It is the customer's responsibility to validate that a particular product with the properties described in the product specification is suitable for use in a particular application. Parameters provided in datasheets and / or specifications may vary in different applications and performance may vary overtime. All operating parameters, including typical parameters, must be validated for each customer application by the customer's technical experts. |