The Dytran model 4010 is a microprocessor controlled, 3-channel DC signal conditioner
amplifier designed for use with bridge-type or differential output accelerometers and
pressure sensors for a variety of general purpose laboratory testing applications. Units
incorporate a variable gain adjustment, shunt calibration capability and multiple
excitation level settings.
A microprocessor SLEEP mode is utilized to eliminate high frequency clock noise and
their associated harmonics. The microprocessor WAKES momentarily to acknowledge
front panel switch depressions then goes to SLEEP immediately after processing and
executing the requested function. This allows the amplifiers to operate within minimum
self generated noise and provides clean, clock free signals.
The Dytran model 4010 incorporates dual 12-bit DAC’s for each channel, to auto-zero the
input and output amplifiers for DC input signals. Input signals with magnitudes of ±10
VDC can be zeroed. A unique output DAC trimming routine allows trimming the output
zero within ±1 mVDC.
The transducer excitation supplies are individually adjustable for each channel from 0.00
to 12.00 VDC. Any setting above 12.00 VDC will generate an excitation voltage of 12.10
VDC. The outputs are short circuit protected and can supply up to 30 mA each. Remote
sense leads are provided to eliminate errors caused be long cable lengths.