COMPANY HEADER
Request Information
  • Ultra Low Frequency Response
  • Ultra Low Noise Level
Unit of Measure

Specifications

Brands

N/A Dytran

Physical

Housing Material

N/A Ti-6 AL-4V

Maximum Weight

N/A 0.14 oz4.0 g

Mounting

N/A Adhesive

Connector Type

N/A 4-Pin

Note for Connector Type

N/A M4.5 X 0.35 Connector. Mates with Dytran cable Model 6893AXX.

Connector Material

N/A Ti-6 AL-4V

Sensing Element Material

N/A Ceramic

Sensing Element Mode

N/A Shear

Environmental

Maximum Peak Shock

N/A 5000 g49050 m/s²

Maximum Peak Vibration (±)

N/A 3000 g29430 m/s²

Environmental Seal

N/A Hermetic

Transducer Electronic Data Sheet (TEDS) Operating Temperature

N/A -40 to +185 ºF-40 to +85 ºC

Base Strain Sensitivity

N/A 0.06 g/µε0.59 m/s²/µε

Applications

Applications

N/A General purpose triaxial vibration measurements Modal and structural analysis Noise, Vibration and Harshness (NVH) Squeak & Rattle Vibration control

Performance

Note for Sensitivity

N/A Measured at 100 Hz, 1 Grms per ISA RP 37.2.

Resonant Frequency

N/A >24 kHz

Full Scale (F.S.) Linearity (±)

N/A 1 %

Note for Full Scale (F.S.) Linearity (±)

N/A Measure using zero-based straight line method, % of F.S. or any lesser range.

Maximum Transverse Sensitivity

N/A 6 %

Sensitivity +15/-10%

N/A 50 mV/g5.1 mV/m/s²

±5 Volts (V) Peak Acceleration Range

N/A 100 g981 m/s²

Maximum Root-Mean-Square (RMS) Noise Floor

N/A 0.00095 g0.009 m/s²

Root-Mean-Square (RMS) Spectral Noise at 1 Hertz (Hz) Frequency

N/A 207 µg2031 µm/s²

Root-Mean-Square (RMS) Spectral Noise at 10 Hertz (Hz) Frequency

N/A 52 µg510 µm/s²

Root-Mean-Square (RMS) Spectral Noise at 100 Hertz (Hz) Frequency

N/A 32 µg314 µm/s²

Root-Mean-Square (RMS) Spectral Noise at 1000 Hertz (Hz) Frequency

N/A 11 µg108 µm/s²

Root-Mean-Square (RMS) Spectral Noise at 10000 Hertz (Hz) Frequency

N/A 8 µg78 µm/s²

+15/-10 Percent (%) Frequency Range

N/A 0.54 to 10,000 Hz

Electrical

Note for Power Supply Current

N/A Do not apply power to this system without current limiting, 20 mA MAX. To do so will destroy the IC charge amplifier.

Compliance Voltage Range

N/A +18 to +30 V

Direct Current (DC) Operating Bias Voltage

N/A +11 to +13 V

Discharge Time Constant

N/A 0.5 to 1.5 sec

Transducer Electronic Data Sheet (TEDS) Feature

N/A IEEE 1451.4

Typical Output Impedance

N/A 100 O

Power Supply Current Range

N/A 2 to 20 mA

Case Isolation

N/A 10 GO

Supplied Accessories

Supplied Accessories

N/A
  1. Accredited calibration certificate (ISO 17025)

Features

Features

N/A Case isolated version available Extended low frequency response Hermetic Low noise JFET electronic Miniature design TEDS options Titanium Triaxial

Note

Note

N/A In the interest of constant product improvement, we reserve the right to change specifications without notice. It is the customer's responsibility to validate that a particular product with the properties described in the product specification is suitable for use in a particular application. Parameters provided in datasheets and/or specifications may vary in different applications and performance may vary overtime. All operating parameters, including typical parameters, must be validated for each customer application by the customer's technical experts. Units on the line drawing are in inches, units in brackets are in millimeters.